Chapter 7 Performance Test
7-26
7.3.2
Amplitude accuracy
(1) Specification
(–2%–offset amplitude accuracy) to (+2%+offset amplitude accuracy)
The offset amplitude accuracy after calibration is as shown in the
following figure.
Figure 7.3.2-1 Offset Amplitude Accuracy
(2) Setup
Figure 7.3.2-2 Amplitude Accuracy Test Connection Diagram
(When using a PPG of MP2110A)
0
5
10
15
20
25
0
50
100
150
200
250
300
350
400
A
m
plitu
de A
cc
u
racy
(
±
mV
)
Difference from Offset Voltage (mV)
Scale=250 mV/div
Scale=200 mV/div
Scale=100 mV/div
Scale=50 mV/div
Scale=15 mV/div
Scale=1 mV/div
Coaxial Cable
+
-
DC Power Supply