Chapter 6 How to Operate Sampling Scope
6-64
Procedure
1. For
MP2110A-054
Connect
the
CRU Out
and
Trigger Clk In
connectors with a
supplied U-link coaxial cable (SMA).
If the Scope module is equipped with
O/E Monitor Out
connector,
Clock Recovery can be executed from Channel B optical signal
input by connecting the
O/E Monitor Out
and
CRU In
connectors
with a U-link coaxial cable (K). If the
O/E Monitor Out
and
CRU
In
connectors are not connected, make sure to install a coaxial
terminator to the
O/E Monitor Out
connector.
For
MP2110A-055
Connect
the
Data Out
connector of 53G CRU and the
SMF In
connector of the scope, using a single-mode optical fiber. Connect
the
Recovered Clock Out
connector of 53G CRU and the
Trigger
Clk In
connector of the scope, using a coaxial cable (SMA).
Trigger Clk In
2Vp-p Max
Ch B In
±
2V Max
Scope
Status Fail
CRU In
1Vp-p Max
CRU Out
U Link Coaxial Cable (SMA)
Trigger Clk In
2Vp-p Max
Ch B In
Scope
Status Fail
MMF
+5dBm Peak Max
CRU In
1Vp-p Max
CRU Out
O/E Monitor Out
U Link Coaxial Cable (K)
U Link Coaxial Cable (SMA)
Trigger Clk In
2Vp-p Max
Scope
Status Fail
U Link Coaxial Cable (SMA)
CRU
Status
Fail
Data Out
Data In
Optical
SMF
Ch B In
MMF
SMF
+5dBm Peak Max
To input the measured signal
Summary of Contents for BERTWave Series
Page 26: ...VI...
Page 74: ...Chapter 1 Outline 1 48...
Page 166: ...Chapter 4 Screen Operation 4 24...
Page 210: ...Chapter 6 How to Operate Sampling Scope 6 16 Figure 6 2 2 4 Switching Graph Display...
Page 309: ...6 9 Measuring Waveform 6 115 6 How to Operate Sampling Scope Figure 6 9 5 1 Marker Display...
Page 322: ...Chapter 6 How to Operate Sampling Scope 6 128...
Page 380: ...Chapter 8 Maintenance 8 14 7 Click Reinstall Windows 8 Click Yes...
Page 432: ...Appendix A Specifications A 36...
Page 458: ...Appendix D Performance Test Record Form D 12...
Page 466: ...Index Index 6...