1. Error Display and How to Take a Measure
205
ML-2350A/2351A
Chapter 2 Inspection and Measure T
o Be T
aken at Occurrence of an Error
Introduction Part
Operating Part
Maintenance Part
Appendixes
Installation
and
Preparation
Part
No.
Contents of error
High
voltage
Alarm
output
Measures
14
SIMMER TROUBLE
OFF
ON
Flashlamp is not turned on.
Check the flashlamp for malfunction.
If faulty, replace it. Resistivity of the
secondary cooling water may below.
Referring to 13, take measures for the
secondary cooling water.
15
CHARGE TROUBLE
OFF
ON
Too much time is required to charge the
capacitor in the laser power supply or the
charged voltage is too high.
Check if the capacity of the power supply
or power cable is insufficient.
Check the fuse in the power supply.
If it is blown, replace it.
16
BANK ALARM
OFF
ON
This signal is output when a trouble of
undercharge, over-charge, no-voltage
or over-voltage of capacitor in the power
supply of the Laser occurs. In this case,
consult us.
18
MAIN SHUTTER TROUBLE
OFF
ON
A laser start signal is input while the
resonator shutter is moving.
Extend the time from input of resonator
shutter operating signal to input of laser
start signal. If the trouble will continue even
after the above measures, consult us.
19
20
21
22
23
24
BRANCH SHUTTER 1 TROUBLE
BRANCH SHUTTER 2 TROUBLE
BRANCH SHUTTER 3 TROUBLE
BRANCH SHUTTER 4 TROUBLE
BRANCH SHUTTER 5 TROUBLE
BRANCH SHUTTER 6 TROUBLE
OFF
ON
A laser start signal is input while the branch
shutter is operating.
Extend the time from input of beam select
signal to input of laser start signal. If the
trouble will continue even after the above
measures, consult us.
28
DISCHARGE UNIT
OVERCURRENT
OFF
ON
Lamp or discharge unit (power transistor
component, CT) fault.
If trouble will continue even after the lamp
has been replaced, consult us.
29
DISCHARGE UNIT TEMP
OFF
ON
Check discharge unit for absence of the
secondary cooling water; and fan motor on
the rear for not running.
30
DISCHARGE UNIT OVERPOWER
OFF
ON
Laser oscillation efficiency lowered.
This may be due to the expired service
life of the lamp, deviation of oscillation,
contaminated YAG rod or mirror, or
failure of the sensor unit. If the trouble
will continue even after lamp has been
replaced, consult us.
31
BRANCH UNIT COVER OPENED
OFF
ON
The branch unit cover plate is removed, or
its mounting screws are loosened.
Secure the cover plate with the screws.
Summary of Contents for ML-2350A-CE
Page 1: ...AA06OM1173714 18 OPERATION MANUAL YAG LASER WELDER ML CE 2350A 2351A Original instructions...
Page 19: ...Introduction Part...
Page 20: ......
Page 37: ...Installation and Preparation Part...
Page 38: ......
Page 46: ......
Page 57: ...Operating Part...
Page 58: ......
Page 118: ......
Page 128: ......
Page 150: ......
Page 174: ......
Page 179: ...Maintenance Part...
Page 180: ......
Page 209: ...Appendixes...
Page 210: ......