140
ML-2350A/2351A
3. Connector Functions
Emergency Stop for Complicated Systems
Interfacing with External Emergency Stop Circuitry
User Supplied Safety Relay Module(s) Required
Complicated Systems are those in which more than one emergency stop sub-circuit
must be linked together. An example of this would be a machine that has a Laser,
parts handler with pneumatic controls, PLC, and conveyor belt all of which have
ESTOP buttons where one ESTOP button stops all devices. Any situation in which
more than one device must be connected together and respond identically to an
emergency stop event is considered a complicated system.
Complicated systems are integrated using certified safety controllers or safety relays.
In this situation one device is the “master” and the rest of the devices are the “slaves”.
The Laser is considered a slave device in this configuration and its emergency
stop must be controlled by the larger machine’s safety controller. The output of the
external safety relay module closes the input to the Laser safety unit and allows the
system to clear the emergency stop state.
ES
TO
P
E
ST
O
P
ESTOP CH2 rtn
ESTOP CH 2
ESTOP CH 1
ESTOP Out 1a
ESTOP Out 1b
ESTOP Out 2a
ESTOP Out 2b
Inside the laser
User ESTOP Wiring
6
2
8
1
3
7
4
5
9
12
10
11
14
15
16
ESTOP CH1 rtn
Interlock rtn 2
Interlock out 2
Interlock rtn 1
Interlock out 1
User Interlock Wiring
(dry contact closure)
17
EXT.I/O (4) Connector
Summary of Contents for ML-2350A-CE
Page 1: ...AA06OM1173714 18 OPERATION MANUAL YAG LASER WELDER ML CE 2350A 2351A Original instructions...
Page 19: ...Introduction Part...
Page 20: ......
Page 37: ...Installation and Preparation Part...
Page 38: ......
Page 46: ......
Page 57: ...Operating Part...
Page 58: ......
Page 118: ......
Page 128: ......
Page 150: ......
Page 174: ......
Page 179: ...Maintenance Part...
Page 180: ......
Page 209: ...Appendixes...
Page 210: ......