2. Maintenance of the Cooler Unit Section
2. Maintenance of the Cooler Unit Section
This section explains the maintenance of the coolant unit, including cleaning of
the solenoid valve filter, water drain of the primary cooling water and secondary
cooling water tanks, maintenance of the ion-exchanger, and water drain for the
case where the laser welder is not used for a long time.
Cleaning the Solenoid Valve Filter
The solenoid valve has a function to adjust the quantity of primary cooling water. If
the solenoid valve is clogged with dust, the primary cooling water is hard to flow,
thereby causing overheating. Perform cleaning for the solenoid valve once every
year as standard.
Item required
Brush, screwdriver, and bucket (container to receive water)
Operating Procedure
(1) Remove the solenoid valve cover
on the rear surface of the main unit
and pull out the solenoid valve.
(2) Turn off the primary cooling water
and disconnect the hose from the
cooling water inlet, and drain the
water.
⇒
Take care not to splash over the
Laser as water may spurt. If
water has splashed, wipe it off
completely.
(3) Remove the 4 body mounting
screws of the solenoid valve and
then remove the upper part of the
solenoid valve.
The part between the staffing
assembly and the gasket is
removed from the main body.
Staffing assembly
Main valve screw
Diaphragm assembly
Solenoid valve cover
Pull out the solenoid valve
Solenoid valve
Rear of the
Laser
Valve body
Gasket
Main body
Body mounting screw
(4 pieces)
Summary of Contents for ML-2350A-CE
Page 1: ...AA06OM1173714 18 OPERATION MANUAL YAG LASER WELDER ML CE 2350A 2351A Original instructions...
Page 19: ...Introduction Part...
Page 20: ......
Page 37: ...Installation and Preparation Part...
Page 38: ......
Page 46: ......
Page 57: ...Operating Part...
Page 58: ......
Page 118: ......
Page 128: ......
Page 150: ......
Page 174: ......
Page 179: ...Maintenance Part...
Page 180: ......
Page 209: ...Appendixes...
Page 210: ......