Commands in DSR Application Subsystems
Output Parameter Commands
276
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:OUTPut:DIMPedance:EXTernal
Syntax
SGENeral:
PDATa(*)
:OUTPut:DIMPedance:EXTernal <Impedance>
SGENeral:
PDATa(*):TERMinal(*)
:OUTPut:DIMPedance:EXTernal
<Impedance>
SGENeral:
PPULse(*)
:OUTPut:DIMPedance:EXTernal <Impedance>
SGENeral:
PPULse(*):TERMinal(*)
:OUTPut:DIMPedance:EXTernal
<Impedance>
CGR(*):MOD(*):
CONNector(*)
:OUTP:DIMP:EXT <Impedance>
Parameters
<Impedance>
Sets the impedance between the OUT and \OUT\
output frontend when the differential termination model is selected
(see
“:OUTPut:TCONfig” on page 274
).
Usually 100 Ohms are used for termination.
Example
:_test:SGEN:PDAT:OUTP:DIMP:EXT 100
:_test:SGEN:PDAT1:TERM2:OUTP:DIMP:EXT 50
:_test:SGEN:PPUL:OUTP:DIMP:EXT 100
:_test:SGEN:PPUL2:TERM3:OUTP:DIMP:EXT 100
:OUTPut:DIMPedance:EXTernal?
Syntax
SGENeral:
PDATa(*)
:OUTPut:DIMPedance:EXTernal?
SGENeral:
PDATa(*):TERMinal(*)
:OUTPut:DIMPedance:EXTernal?
SGENeral:
PPULse(*)
:OUTPut:DIMPedance:EXTernal?
SGENeral:
PPULse(*):TERMinal(*)
:OUTPut:DIMPedance:EXTernal?
CGRoup(*):MODule(*):
CONNector(*)
:OUTPut:DIMPedance:EXTernal?
Return Value
Returns the differential termination resistor (see
“:OUTPut:TCONfig”
on page 274
).
Example
:_test:SGEN:PDAT:OUTP:DIMP:EXT?
:_test:SGEN:PDAT1:TERM2:OUTP:DIMP:EXT?
:_test:SGEN:PPUL:OUTP:DIMP:EXT?
:_test:SGEN:PPUL2:TERM3:OUTP:DIMP:EXT?
:_test:MOD2:CONN3:OUTP:DIMP:EXT?
might return:
100