Timing Parameter Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
189
:PULSe:CROSsing
Syntax
:<Handle>:SGENeral:PDATa(*):[:SOURce]:PULSe:CROSsing <CROSsing>
:<Handle>:SGENeral:PDATa(*):TERMinal(*)[:SOURce]:PULSe:CROSsing
<CROSsing>
:<Handle>:SGENeral:PPULse(*):[:SOURce]:PULSe:CROSsing <CROSsing>
:<Handle>:SGENeral:PPULse(*):TERMinal(*)[:SOURce]:PULSe:CROSsing
<CROSsing>
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*)[:SOURce]:PULSe:CROSsing
<CROSsing>
Parameters
<CROSsing>
Cross point of the signal in percentage of signal
amplitude. The formula for the cross point is:
Cross Point = 100 * ( (V
Cross
– V
High
) / (V
High
– V
Low
) )
This parameter is valid only for NRZ format modes.
Cross point range is subject to change; depending on hardware
capabilities.
Valid range for E4862B: 25 ... 75 %
Valid range for N4872A/74A: 20 ... 80 %
N O T E
This parameter can also be used for the N4868A 10.8G Booster
modules. The valid range is: 40 ... 60 %
Example
:_test:SGEN:PDAT1:PULS:CROS 40.0
:_test:SGEN:PDAT1:TERM1:SOUR:PULS:CROS 40.0
:_test:SGEN:PPUL1:PULS:CROS 40
:_test:SGEN:PPUL1:TERM1:SOUR:PULS:CROS 40
:_test:MODULE2:CONNECTOR3:PULSE:CROSSING 40.0