Commands in DSR Application Subsystems
Amplitude/Offset Correction Factor Commands
236
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Example
:_test:SGEN:PDAT1:CAL:OCF?
might return:
8.0E–1
:CALibration[:VOLTage]:ACFactor
Syntax
:<Handle>:CGRoup(*):MODule(*):CONNector(*):CALibration[:VOLTage]:
ACFactor <acfactor>
:<Handle>:CGRoup(*):MODule(*):CONNector(*):CLOCk:CALibration[:
VOLTage]:ACFactor <acfactor>
:<Handle>:SGENeral:PDATa(*):CALibration[:VOLTage]:ACFactor <acfactor>
:<Handle>:SGENeral:PDATa(*):CLOCk:CALibration[:VOLTage]:ACFactor
<acfactor>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):CALibration[:VOLTage]:ACFactor
<acfactor>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):CLOCk:CALibration[:VOLTage]:
ACFactor <acfactor>
Parameters
<acfactor>
Amplitude correction factor; this can be in the range
0.0 ... 1.0. See
“Amplitude/Offset Correction Factor Commands” on
page 234
for details of this command.
Example
:_test:SGEN:PDAT1:CAL:ACF 0.75
:CALibration[:VOLTage]:ACFactor?
Syntax
:<Handle>:CGRoup(*):MODule(*):CONNector(*):CALibration[:VOLTage]:
ACFactor?
:<Handle>:CGRoup(*):MODule(*):CONNector(*):CLOCk:CALibration[:
VOLTage]:ACFactor?
:<Handle>:SGENeral:PDATa(*):CALibration[:VOLTage]:ACFactor?
:<Handle>:SGENeral:PDATa(*):CLOCk:CALibration[:VOLTage]:ACFactor?
:<Handle>:SGENeral:PDATa(*):TERMinal(*):CALibration[:VOLTage]:
ACFactor?
:<Handle>:SGENeral:PDATa(*):TERMinal(*):CLOCk:CALibration[:VOLTage]:
ACFactor?
Return Value
Returns the current amplitude calibration factor. See
“Amplitude/Offset Correction Factor Commands” on page 234
for
details of this command.