Commands in DSR Application Subsystems
Port Administration Commands
300
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Port Administration Commands
The following commands are available for managing pulse and data
ports. The parameters can be specified separately for pulse and data
ports.
The MUX commands—also available for pulse and data ports—are
described in
“Timing Parameter Commands” on page 184
.
:APPend
Syntax
:<Handle>:SGENeral:
PDATa(*):
APPend <“Port Type”> [, <No. of Terminals>]
[, <“Port Name”>], [<Species>]
:<Handle>:SGENeral:
PPULse(*):
APPend <“Port Type”> [, <No. of Terminals>]
[, <“Port Name”>], [<Species>]
Parameters
<“Port Type”>
A quoted string of the port type, either
INPUT_PORT or OUTPUT_PORT, seen from DUT point of view.
[<No. of Terminals>]
The integer number of terminals used in the
new port (<NR1>). This is an optional parameter.
[<“Port Name”>]
A quoted string of the name used to identify the
port. This is an optional parameter.
[<Species>]
(Optional) Species of the port. Possible values are
electrical (ELEC) or optical (OPT). All terminals of the port are
assigned this species. Default is electrical.
Creates and appends a new port to the list of ports. The port type is
required. The number of terminals, the port name, and species are
optional. The suffix of PDATa and PPULse is ignored.
Example
:_test:SGEN:PDAT:APP "INPUT_PORT", 4, "DataBus", ELEC
:_test:SGEN:PPUL:APP "INPUT_PORT", 4, "DataBus", ELEC
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com