Parameter Commands for MUX and DEMUX Modules
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
209
:OUTPut:EATTenuator
Syntax
:<Handle>[:CGRoup*]:MODule*:CONNector*:OUTPut:EATTenuator <value>
Description
This command allows you to enter the attenuation (in dB) of an
external attenuator on the multiplexer’s output. When this value is
changed, the output values are also changed so that they reflect the
values that actually reach the DUT (and not the values that reach the
attenuator).
Related Commands
•
“:VOLTage:AMPLitude” on page 206
•
“:VOLTage:OFFSet” on page 207
•
“:VOLTage:OFFSet:ENABle” on page 208
Parameters
<value>
Attenuation value of the external attenuator in dB; the
range is dependent on the offset voltage.
Example
:_test:_test:MOD10:CONN1:OUTP:EATT 20
:OUTPut:EATTenuator?
Syntax
:<Handle>[:CGRoup*]:MODule*:CONNector*:OUTPut:EATTenuator?
Return Value
Returns the attenuation value that was entered (manually or by the
program) for the external attenuator in dB; range: 0 – 120 dB.
Example
:_test:MOD10:CONN1:OUTP:EATT?
might return:
2.0E+1