:SGENeral:GLOBal Subsystem
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
143
:<Handle>:SGENeral:GLOBal:CALibration:
PUSe PASSive
Syntax
:<Handle>:SGENeral:GLOBal:CALibration:PUSe PASSive
Description
Instead of assuming an active probe, this option uses the customer-
provided levels. This is useful when you use long cables, for example
for testing the pins of an integrated circuit.
Example
:_test:SGEN:GLOB:CAL:PUS PASS
:<Handle>:SGENeral:GLOBal:CALibration:
SELF DELay
Syntax
:<Handle>:SGENeral:GLOBal:CALibration:SELF DELay
Description
Any time frontends were exchanged in a module, the system must be
autocalibrated after a 30-minute warmup period. Delay Auto
Calibration must be performed before Zero Adjustment or Cable
Adjustment.
Related Commands
•
“:CALibration:CDELay” on page 303
•
“:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):CALibration:
ZDELay” on page 131
Example
:_test:SGEN:GLOB:CAL:SELF DEL
:<Handle>:SGENeral:GLOBal:DOFFset
Syntax
:<Handle>:SGENeral:GLOBal:DOFFset <DelayOffset>
Parameters
<DelayOffset>
A numeric value out of the specified range, for
example, 10e-9.
This command specifies an offset value to the fixed delay for all
connectors. So, it is possible to set negative delays for individual
connectors to achieve setup and hold times behavior.
Example
:_test:SGEN:GLOB:DOFF 10e-9