Commands in DSR Application Subsystems
Output Parameter Commands
274
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:OUTPut:TCONfig
Syntax
:<Handle>:SGENeral:
PDATa(*):OUTPut:
TCONfig <VOLTage | DIFFerential>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):OUTPut:
TCONfig
<VOLTage | DIFFerential>
:<Handle>:SGENeral:
PPULse(*):OUTPut:
TCONfig <VOLTage | DIFFerential>
:<Handle>:SGENeral:
PPULse(*):TERMinal(*):OUTPut:
TCONfig
<VOLTage | DIFFerential>
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):OUTPut:
TCONfig
<VOLTage | DIFFerential>
Parameters
<VOLTage | DIFFerential>
Selects the termination model for
output frontends:
• VOLTage selects termination via resistor against a termination
voltage (this is the reset value).
If VOLTage is selected, the termination resistor and termination
voltage can be set with
“:OUTPut:IMPedance:EXTernal” on
page 273
and
“:OUTPut:TVOLtage” on page 272
.
• DIFFerential selects termination via a resistor against the
complement output.
If DIFFerential is selected, the differential termination resistor can
be set with
“:OUTPut:DIMPedance:EXTernal” on page 276
.
Description
For output frontends, it is necessary to specify how the signal will be
terminated. There are two possible termination models available:
• Via a resistor against a termination voltage; the resistor might be
high impedance. Optimum signal performance is achieved with 50
Ohm. In the following picture, this is shown for single ended and
differential outputs.
50 / HiZ
V
OUT
50
50
V
OUT
~OUT
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com