![Agilent Technologies 81250 System Programming Manual And Scpi Reference Download Page 96](http://html.mh-extra.com/html/agilent-technologies/81250/81250_system-programming-manual-and-scpi-reference_2868601096.webp)
:MMEMory Subsystem
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
95
:<Handle>:MMEMory:SETTing:LOAD
Syntax
:<Handle>:MMEMory:SETTing:LOAD [<“SettingName”>] [,<YDELay | NDELay
| ODELay>]
Parameters
Optional first parameter
[<“SettingName”>]
If not specified, the current setting is used. The
setting at system power-up is “Default”. If changes are stored with
‘save-as’, the current setting could be “TEST1”.
YDELay
Yes delay; load setting with cable delay parameters. This is
the default.
NDELay
No delay; load setting without cable delay parameters.
ODELay
Only delay; load only cable delay parameters.
Description
The command is used to load the current setting or a specified setting
into the Agilent 81250 System’s database.
Example
:_test:MMEM:SETT:LOAD "TEST1"
:<Handle>:MMEMory:SETTing:SAVE
Syntax
:<Handle>:MMEMory:SETTing:SAVE [<“SettingName”>]
Parameters
[<“SettingName”>]
If not specified, the changes are saved into the
current setting. After system power-up, the name of the current setting
is “Default”.
Description
The command is used to save into either the current setting or into a
new setting.
Return Value
None
Example
:_test:MMEM:SETT:SAVE "TEST"
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com