Optical Output Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
281
The following table shows the relationship between the
:OUTPut:OPTic:
METHod
parameter and the commands for setting the power levels:
The parameters are calculated as follows (based on the High and Low
power values). Note that all powers are given in Watt.
• Average Power = (High + Low)/2
• Modulation Amplitude = High - Low
• Extinction Ratio
None
= High/Low
• Extinction Ratio
dB
= 10*log(High/Low)
See also
“Optical Input Commands” on page 263
.
:OUTPut:OPTic[:STATe][:VALue]
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):OUTPut:OPTic:[STATe]:
[VALue] <OFF|ON>
:<Handle>:SGENeral:PDATa(*):OUTPut:OPTic:[STATe]:[VALue] <OFF|ON>
:<Handle>:SGENeral:PPULse(*):OUTPut:OPTic:[STATe]:[VALue] <OFF|ON>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):OUTPut:OPTic:[STATe]:[VALue]
<OFF|ON>
:<Handle>:SGENeral:PPULse(*):TERMinal(*):OUTPut:OPTic:[STATe]:[VALue]
<OFF|ON>
Parameters
<OFF|ON>
Enables/disables the laser on the E/O converter. The
electrical board is enabled/disabled automatically if the connector
species is set to OPT.
Examples
:_test:CGR1:MOD1:CONN1:OUTP:OPT:STAT:VAL OFF
:_test:CGR1:MOD1:CONN1:OUTP:OPT OFF
Parameters
:OUTPut:OPTic:METHod
AVER_ERAT
AVER_MAMP
HIGH_LOW
High
Calculated
Calculated
Direct entry
Low
Calculated
Calculated
Direct entry
Average
Direct entry
Direct entry
Calculated
Modulation
Amplitude
Calculated
Direct entry
Calculated
Extinction
Ratio
Direct entry
Calculated
Calculated