Commands in DSR Application Subsystems
:SGENeral:GLOBal:CONFigure Subsystem
152
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:<Handle>:SGENeral:GLOBal:CONFigure[:
ECOunt]
Syntax
:<Handle>:SGENeral:GLOBal:CONFigure[:ECOunt][FAILures | ONESfailed |
ZERosfailed]
In the Error Rate (Error COunt) measurement mode, an expected data
segment (or more) is compared and the failed bits are counted. To read
the results, see
“:FETCh[:ECOunt]?” on page 312
.
If the argument is omitted, the default mode is count all FAILures.
Example
:_test:SGEN:GLOB:CONF ONES
:<Handle>:SGENeral:GLOBal:CONFigure:
ECAPture
Syntax
:<Handle>:SGENeral:GLOBal:CONFigure:ECAPture[<StopBits>]
Range/Value Coupling
On E4832A and E4861A modules, there is no restriction for the
frequency multiplier.
The frequency multiplier can be queried via CGR:MOD:CONN:MUX?
The segment resolution can be queried via SGEN:GLOB:MUX?
Description
In the Error CAPture mode, the incoming data stream is compared to a
stream of expected data. After compare, the resulting data stream is
held in the internal memory ready to be displayed or to be saved to the
database. An additional parameter can be specified called <StopBits>.
This is a number of cycles that can be specified to stop the analyzer
after an error occurs. If the parameters are omitted following default
values are assumed: StopBits= 32768.
For systems housing E4832A or E4861A modules, the minimum
number of stop bits is 976.
Note that the value is always rounded up internally because of the
resolution. Note that because of clock uncertainty, a few additional
bits (depending on resolution) are recorded.
Example
:_test:SGEN:GLOB:CONF:ECAP 1024