Commands in DSR Application Subsystems
Parameter Commands for MUX and DEMUX Modules
210
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:OUTPut:DELay:SWEep
Syntax
:<Handle>[:CGRoup*]:MODule*:CONNector*:OUTPut:DELay:SWEep <sweep>
Description
Manipulates globally the delay for all generators bound to the inputs of
the E4868A module. The values set with this command are added to
the start delay (see
“:PULSe:DELay” on page 185
) of those generators.
Parameters
<sweep>
The valid values depend on the current delay mode:
• Delay correction mode (
TCOM
): -200 ps to +200 ps.
• Delay calibration mode (
DCAL
): -2 ns to +2 ns
The current mode can be queried with
“:OUTPut:DELay:SWEep:
MODe?” on page 212
.
Example
In delay correction mode, set the delay to 3 ps:
:_test:MOD10:CONN1:OUTP:DEL:SWE 3E-12
In delay calibration mode, set the delay to -100 ps:
:_test:MOD10:CONN1:OUTP:DEL:SWE -100E-12