Input Parameter Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
247
:INPut:TYPE
Syntax
:<Handle>:SGENeral:
PDATa(*)
:INPut:TYPE <BALanced | NORMal | INVerted>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*)
:INPut:TYPE <BALanced |
NORMal | INVerted>
:<Handle>:CGRoup(*):MODule(*):
CONNector(*)
:INPut:TYPE <BALanced |
NORMal | INVerted>
Parameters
<BALanced | NORMal | INVerted>
Selects how the differential
input frontends are sampled.
• BALanced selects differential operation (this is the reset value).
• NORMal compares the normal input against a threshold value.
• INVerted compares the complement input against a threshold value.
In NORMal and INVerted operation (
“:INPut:IMPedance[:INTernal]”
on page 252
and
“:INPut:TVOLtage” on page 250
) specify the
termination of the input.
In BALanced operation the termination is not used, but is checked
against absolute limits.
“:INPut:SERial” on page 254
is also ignored in
BALanced operation, but is checked against absolute limits.
Example
:_test:SGEN:PDAT1:INP:TYPE BAL
:_test:SGEN:PDAT1:TERM1:INP:TYPE INV
:_test:CGROUP1:MODULE2:CONNECTOR4:INPut:TYPE NORM
V
T
R
T
=50
Ω
R
T
=50
Ω
R
T
=100
Ω
V
Thres
V
Thres
Termination
D
iffe
re
ntia
l
Cent
er t
apped
Sampling
Nor
m
al
V
in
> V
th
re
s
= 1
In
In
+
-
+
-
+
-
In
ve
rte
d
V
in
> V
th
re
s
= 0
Di
ff
er
ent
ial
V
in
> V
in
=1
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com