:SGENeral:GLOBal:TRIGger Subsystem
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
173
:SGENeral:GLOBal:TRIGger
Subsystem
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer]
This command subsystem is used to set and measure the clock and
clock reference.
:<Handle>:SGENeral:GLOBal:TRIGger[:
SEQuence][:LAYer][:SOURce]
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer][:SOURce]
<CDR | INT10 | VXI10 | EXT1 | EXT2 | EXT5 | EXT10 | EXTernal|
EXTDirect>
Description
This command sets the clock/reference or the external clock mode.
The arguments available are:
For more information on clock sources, please refer to the
Agilent
81250 Parallel Bit Error Ratio Tester System User Guide.
N O T E
If the external clock mode is set and the external start mode is
selected, the external clock must be in a stable state for 100 ms before
the external start can take place.
Example
:_test:SGEN:GLOB:TRIG EXT5
CDR
Clock Data Recovery (only available for E4809A)
INT10 reference
internal clock source, 10 MHz internal clock reference
VXI10 reference
internal clock source, 10 MHz VXI backplane reference
EXT1 reference
internal clock source, 1 MHz external clock reference
EXT2 reference
internal clock source, 2 MHz external clock reference
EXT5 reference
internal clock source, 5 MHz external clock reference
EXT10 reference
internal clock source, 10 MHz external clock reference
EXTernal
external clock source
EXTernal Direct
external clock source direct, not handled by ParBERT
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com