Commands in DSR Application Subsystems
Input Parameter Commands
262
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:INPut:DIMPedance?
Syntax
:<Handle>:SGENeral:
PDATa(*):INPut
:DIMPedance[:INTernal]?
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):INPut
:DIMPedance[:INTernal]?
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):INPut
:DIMPedance[:
INTernal]?
Return Value
Returns the differential termination resistor (see
“:INPut:TCONfig” on
page 260
).
The only possible value is 100 Ohm.
Example
:_test:SGEN:PDAT:INP:DIMP?
:_test:SGEN:PDAT1:TERM1:INP:DIMP?
:_test:CGROUP1:MODULE2:CONNECTOR4:INPut:DIMP?
might return:
100
:INPut:DISConnect:MODE?
Syntax
:<Handle>:CGRoup(*):MODule(*):CONNector(*):INPut:DISConnect:MODE?
:<Handle>:SGENeral:PDATa(*):INPut:DISConnect:MODE?
:<Handle>:SGENeral:PDATa(*):TERMinal:INPut:DISConnect:MODE?
Return Value
Returns how the specified connector is to be disconnected when the
SGENeral:GLOBal:CONNect OFF
command is executed. Can be either
RELay (disconnected by switching off the corresponding relay) or
VOLtage (disconnected by reducing the voltage to 0).
Example
:_test:CGR1:MOD1:CONN1:INP:DISC:MODE?
might return:
REL