Commands in DSR Application Subsystems
Input Parameter Commands
250
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:INPut:TVOLtage
Syntax
:<Handle>:SGENeral:
PDATa(*):INPut
:TVOLtage <TerminationVoltage>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):INPut
:TVOLtage
<TerminationVoltage>
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):INPut
:TVOLtage
<TerminationVoltage>
Parameters
<TerminationVoltage>
Termination voltage (<NRf>).
Specifies the termination voltage of an analyzer input connector. The
termination voltage range of the available analyzer frontends is –2.1 V
to +3.1 V.
Example
:_test:SGEN:PDAT1:INP:TVOL -2
:_test:SGEN1:PDAT1:TERM1:INP:TVOL -2
:_test:CGROUP1:MODULE2:CONNECTOR4:INP:TVOL -2
:INPut:TVOLtage?
Syntax
:<Handle>:SGENeral:
PDATa(*):INPut
:TVOLtage?
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):INPut
:TVOLtage?
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):INPut
:TVOLtage?
Return Value
Returns the current termination voltage of the specified analyzer
input connector.
Example
:_test:SGEN:PDAT1:INP:TVOL?
:_test:SGEN1:PDAT1:TERM1:INP:TVOL?
:_test:CGROUP1:MODULE2:CONNECTOR4:INP:TVOL?
might return:
-2.000