Level Parameter Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
233
:SENSe:VOLTage:RANGe
Syntax
:<Handle>:SGENeral:
PDATa(*)
:SENSe:VOLTage:RANGe <range>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*)
:SENSe:VOLTage:RANGe <range>
:<Handle>:CGRoup(*):MODule(*):
CONNector(*)
:SENSe:VOLTage:RANGe
<range>
Parameters
<range>
Selects the allowed input voltage range. For E4835A
differential input frontends, two values are supported:
• 3 – allows input voltages –2 … 3 V
• 5 – allows input voltages 0 ... 5 V
For E4863A, E4863B, E4865A, and E4867A frontends, the following
values are supported:
• 0 – allows input voltages –2 … 0 V
• 1 – allows input voltages –1 … 1 V
• 2 – allows input voltages 0 … 2 V
For N4873A and N4875A frontends, the selected value is the upper
value of a 2 V range. For example, 1.5 defines a range of –0.5 ... 1.5 V.
Example
:_test:SGEN:PDAT1:INP:RANG 3
:_test:SGEN:PDAT1:TERM1:SENS:RANG 5
:_test:CGROUP1:MODULE2:CONNECTOR4:SENSE:VOLTAGE:RANGE 3
:SENSe:VOLTage:RANGe?
Syntax
:<Handle>:SGENeral:
PDATa(*)
:SENSe:VOLTage:RANGe?
:<Handle>:SGENeral:
PDATa(*):TERMinal(*)
:SENSe:VOLTage:RANGe?
:<Handle>:CGRoup(*):MODule(*):
CONNector(*)
:SENSe:VOLTage:RANGe?
Return Value
Returns the currently selected input voltage range. See
“:SENSe:
VOLTage:RANGe” on page 233
for details.
Example
:_test:SGEN:PDAT1:SENS:VOLT:RANG?
:_test:SGEN:PDAT1:TERM1:SENS:VOLT:RANG?
:_test:CGROUP1:MODULE2:CONNECTOR4:SENSE:VOLTAGE:RANGE?
might return:
3.0E+0
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com