Commands in DSR Application Subsystems
Parameter Commands for MUX and DEMUX Modules
214
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:INPut:OCOMpensation
N O T E
Valid only for E4869A/B (DEMUX) modules.
Syntax
:<Handle>[:CGRoup*]:MODule*:CONNector*:INPut:OCOMpensation <value>
Definition
By default, the decision threshold for sampling the incoming signal is
set according to the mean value of the data signal. This is adequate for
pure PRBS data. In cases where the received pattern has an
unbalanced ratio of zeros to ones, this command allows you to fine-
tune the threshold to an appropriate value.
An inadequate offset can be suspected, if:
• None of the analyzer terminals could synchronize.
• All analyzer terminals did synchronize, but exhibit an abnormally
high BER.
• Some analyzer terminals did synchronize but show a BER close to
the synchronization threshold, while others did not synchronize at
all.
See also “
How to Change the Input Parameters of a DEMUX Module”
in the
System User Guide
for further details.
Parameters
<value>
New value of offset compensation; range: ±1 (corresponds
to roughly ±100 mV).
Example
:_test:MOD10:CONN1:INP:OCOM 0.9
:INPut:OCOMpensation?
N O T E
Valid only for E4869A/B (DMUX) modules.
Syntax
:<Handle>[:CGRoup*]:MODule*:CONNector*:INPut:OCOMpensation?
Definition
Returns value used for the offset compensation.
Return Value
Current value for offset compensation; range: ±1 (corresponds to
roughly ±100 mV).
Example
:_test:MOD10:CONN1:INP:OCOM?
might return:
9.0E-1