![Agilent Technologies 81250 System Programming Manual And Scpi Reference Download Page 131](http://html.mh-extra.com/html/agilent-technologies/81250/81250_system-programming-manual-and-scpi-reference_2868601131.webp)
Commands in DSR Application Subsystems
[:CGRoup(*)]:MODule(*):CONNector(*) Subsystem
130
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):TYPE?
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):TYPE?
Returns the product number of the frontend to which the connector
belongs. This value is returned in a quoted string.
Example
:_test:CGR1:MOD1:CONN1:TYPE?
might return:
"E4863A"
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):NAME?
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):NAME?
Returns the name of the specified connector as a 7-digit quoted string
according to connector, module and clock group. The following syntax
is used:
ggmmccc
gg
represents the clock group,
mm
represents the module,
ccc
represents the connector.
Example
:_test:CGR1:MOD1:CONN1:NAME?
might return:
"0102001"
The above query returns the name of connector 1 in module 2 of clock
group 1.
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):TNAMe?
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):TNAMe?
Returns the terminal name to which this connector is connected. An
empty “” string is returned if the connector is not connected.
Example
:_test:CGR1:MOD1:CONN1:TNAM?
might return:
"Data1"
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com