Commands in DSR Application Subsystems
Timing Parameter Commands
188
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:PULSe:DELay:TIME
Syntax
:<Handle>:SGENeral:
PDATa(*)
:[:SOURce]:PULSe:DELay:TIME <Time>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*)
:[:SOURce]:PULSe:DELay:TIME
<Time>
:<Handle>:SGENeral:
PPULs(*)
:[:SOURce]:PULSe:DELay:TIME <Time>
:<Handle>:SGENeral:
PPULs (*):TERMinal(*)
:[:SOURce]:PULSe:DELay:TIME
<Time>
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*)
:[:SOURce]:PULSe:DELay:
TIME <Time>
Parameters
<Time>
Delay in seconds.
If the delay is specified by
DELay:TIME
and
PULSe:DELay:TIME
, the
corresponding delay is calculated. The number of cycles is converted
into seconds and added to the specified
PULSe:DELay:TIME
value.
If
PULSe:DELay
is used, the value is interpreted as a timing value only.
This sets
PULSe:DELay:TIME
to zero.
Example
:_test:SGEN:PDAT1:PULS:DEL:TIME 0.5
:_test:SGEN1:PDAT1:TERM1:PULS:DEL:TIME 0.5
:_test:SGEN:PPUL1:PULS:DELAY:TIME 0.5
:_test:CGROUP1:MODULE2:CONNECTOR4:PULS:DELAY:TIME 0.5
:PULSe:DELay:TIME?
Syntax
:<Handle>:SGENeral:
PDATa(*)
:[:SOURce]:PULSe:DELay:TIME?
:<Handle>:SGENeral:
PDATa(*):TERMinal(*)
:[:SOURce]:PULSe:DELay:TIME?
:<Handle>:SGENeral:
PPULs(*)
:[:SOURce]:PULSe:DELay:TIME?
:<Handle>:SGENeral:
PPULs (*):TERMinal(*)
:[:SOURce]:PULSe:DELay:TIME?
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*)
:[:SOURce]:PULSe:DELay:
TIME?
Return Value
Returns the current cycle value of the pulse delay for the specified
generator output connector.
Example
:_test:SGEN:PDAT1:PULS:DELAY:TIME?
:_test:SGEN1:PDAT1:TERM1:PULS:DEL:TIME?
:_test:SGEN:PPUL1:PULS:DEL:TIME?
:_test:CGROUP1:MODULE2:CONNECTOR4:PULSE:DELAY:TIME?
might return:
5.000000E-001