![Agilent Technologies 81250 System Programming Manual And Scpi Reference Download Page 246](http://html.mh-extra.com/html/agilent-technologies/81250/81250_system-programming-manual-and-scpi-reference_2868601246.webp)
Input Parameter Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
245
Input Parameter Commands
The input parameter commands are available on port, terminal and
connector level.
:INPut[:STATe]
Syntax
:<Handle>:SGENeral:
PDATa(*):INPut
[:STATe] <ON | OFF>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):INPut
[:STATe] <ON | OFF>
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):INPut
[:STATe] <ON | OFF>
Parameters
<ON | OFF>
Switch the specified analyzer input connector ON or
OFF.
Example
:_test:SGEN:PDAT1:INPUT ON
:_test:SGEN1:PDAT1:TERM1:INPUT ON
:_test:CGROUP1:MODULE2:CONNECTOR4:INPUT ON
:INPut[:STATe]?
Syntax
:<Handle>:SGENeral:
PDATa(*):INPut
[:STATe]? <NFORce|FORCe>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):INPut
[:STATe]?
<NFORce|FORCe>
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):INPut
[:STATe]?
<NFORce|FORCe>
Description
Returns the current status of the specified analyzer input connector.
Parameters
<NFORce|FORCe>
Defines how the query is executed:
• NFORce
Default value. Only the firmware is queried for the current state.
• FORCe
When FORCe is entered, the firmware checks the current condition
in the hardware. This is required to make sure that the input is
really switched on, or to check if the protection circuit within the
input has turned off the channel to protect it from damage.
N O T E
The
FORCe
call is slower than
NFORce
.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com