:EDIT:SEGMent(*) Subsystem
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
115
:<Handle>:EDIT:SEGMent(*):PATTern:
MODify:INVert
Syntax
:<Handle>:EDIT:SEGMent(*):PATTern:MODify:INVert StartTrace, StartVector,
EndTrace, EndVector
Parameters
StartTrace
Number of start trace.
StartVector
Number of start vector.
EndTrace
Number of end trace.
EndVector
Number of end vector.
(*)
Inspector number of the segment.
Description
Inverts the state information in the bounded area.
This command is only allowed in single-bit codings. Attempting to
execute this command in multi-bit codings causes an error.
Attempting to invert data in a region in an analyzer segment causes an
error.
For example, for the coding “01”, after this command execution, all 0’s
will be changed into 1’s and all the 1’s will be changed into 0’s in the
area.
Example
:_test:EDIT:SEGM3:PATT:MOD:INV 0,0,5,10