Commands in DSR Application Subsystems
[:CGRoup(*)]:MODule(*) Subsystem
126
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
[:CGRoup(*)]:MODule(*)
Subsystem
Syntax
:<Handle>[:CGRoup(*)]:MODule(*)
This tree provides a set of commands which co-operate with a single
module.
:<Handle>[:CGRoup(*)]:MODule(*):
CINFormation?
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CINFormation? <SHORt | DETailed>
Return Value
This query returns the module configuration information, including
the frontends. The response is an expression according to the
following syntax:
The optional syntax elements only appear if DETailed is selected.
Example
:_test:MOD2:CINF? DET
might return for example (formatted here only for readability):
(E4861A,2640,"DE40701397","DE40701397/0",
(E4863A,2640,ANALYZER,"DE40701316","DE40701316/0",S1),
(E4862A,2640,GENERATOR,"DE40701738","DE40701738/0",S1))
<ModuleInfo>
::= “(”<ProductNr> “,” <SpeedClass> [“,” <SerNr> “,”
<IDNr>] “,” <ConnInfo> (“,” <ConnInfo>)* “)”
<ConnInfo>
::= “(”<ProductNr> “,” <SpeedClass> “,” “Type” [“,”
<SerNr> “,” <IDNr> “,” <ConnNr>] “)”
<ProductNr>
::= for example, “E4838A”
<SpeedClass>
::= for example, 660
<Type>
::= “ANALYZER” | “GENERATOR”
<SerNr>
::= serial number (0 if not available)
<IDNr>
::= identification number (0 if not available)
<ConnNr>
::= “S1” | “D1” | “D2”
S1: single frontend
D1: first connector of a dual frontend
D2: second connector of a dual frontend
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com