![Agilent Technologies 81250 System Programming Manual And Scpi Reference Download Page 247](http://html.mh-extra.com/html/agilent-technologies/81250/81250_system-programming-manual-and-scpi-reference_2868601247.webp)
Commands in DSR Application Subsystems
Input Parameter Commands
246
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Example
:_test:SGEN:PDAT1:INPUT?
:_test:SGEN1:PDAT1:TERM1:INPUT?
:_test:CGROUP1:MODULE2:CONNECTOR4:INPUT?
might return:
ON
:INPut:POLarity
Syntax
:<Handle>:SGENeral:
PDATa(*)
:INPut:POLarity <NORMal | INVerted>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*)
:INPut:POLarity <NORMal |
INVerted>
:<Handle>:CGRoup(*):MODule(*):
CONNector(*)
:INPut:POLarity <NORMal |
INVerted>
Parameters
<NORMal | INVerted>
Sets the input polarity of the specified
data
port/terminal/connector to either normal or inverted.
This command is not available for the E4863A/E4865A frontends and
E4867A module.
Example
:_test:SGEN:PDAT1:INP:POL INV
:_test:SGEN:PDAT1:TERM1:INP:POL INV
:_test:CGROUP1:MODULE2:CONNECTOR4:INPut:POLARITY NORM
:INPut:POLarity?
Syntax
:<Handle>:SGENeral:
PDATa(*)
:INPut:POLarity?
:<Handle>:SGENeral:
PDATa(*):TERMinal(*)
:INPut:POLarity?
:<Handle>:CGRoup(*):MODule(*):
CONNector(*)
:INPut:POLarity?
Return Value
Returns the input polarity of the specified port/terminal/connector.
The return value is either NORMal or INVerted.
Example
:_test:SGEN:PDAT1:INP:POL?
:_test:SGEN:PDAT1:TERM1:INP:POL?
:_test:CGROUP1:MODULE2:CONNECTOR4:INPut:POLARITY?
might return:
NORM
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com