![Agilent Technologies 81250 System Programming Manual And Scpi Reference Download Page 284](http://html.mh-extra.com/html/agilent-technologies/81250/81250_system-programming-manual-and-scpi-reference_2868601284.webp)
Optical Output Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
283
:OUTPut:OPTic:METHod[:VALue]
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):OUTPut:OPTic:METHod:
[VALue] <AVER_ERAT | AVER_MAMP | HIGH_LOW>
:<Handle>:SGENeral:PDATa(*):OUTPut:OPTic:METHod:[VALue] <AVER_ERAT
| AVER_MAMP | HIGH_LOW>
:<Handle>:SGENeral:PPULse(*):OUTPut:OPTic:METHod:[VALue]
<AVER_ERAT | AVER_MAMP | HIGH_LOW>
:<Handle>:SGENeral:PDATa(*):TERMinal(*):OUTPut:OPTic:METHod:[VALue]
<AVER_ERAT | AVER_MAMP | HIGH_LOW>
:<Handle>:SGENeral:PPULse(*):TERMinal(*):OUTPut:OPTic:METHod:[VALue]
<AVER_ERAT | AVER_MAMP | HIGH_LOW>
Parameters
<AVER_ERAT | AVER_MAMP | HIGH_LOW>
Sets the method
used to define the optical levels:
•
AVER_ERAT
(Average Power / Extinction Ratio)
•
AVER_MAMP
(Average Power / Modulation Amplitude)
•
HIGH_LOW
(High Power / Low Power)
The optical parameters can always be queried, but only the
parameters specified by
METHod
can be set. For example, if the
METHod
is
AVER_MAMP
, only average power and modulation amplitude can be
directly set. Trying to set high power in this case results in an error.
Examples
:_test:CGR1:MOD1:CONN1:OUTP:OPT:METH HIGH_LOW
:_test:CGR1:MOD1:CONN1:OUTP:OPT:UNIT WATT
:_test:CGR1:MOD1:CONN1:OUTP:OPT:HIGH 0.865242735174E-3
:OUTPut:OPTic:METHod[:VALue]?
Syntax
:<Handle>[:CGRoup(*)]:MODule(*):CONNector(*):OUTPut:OPTic:METHod:
[VALue]?
:<Handle>:SGENeral:PDATa(*):OUTPut:OPTic:METHod:[VALue]?
:<Handle>:SGENeral:PPULse(*):OUTPut:OPTic:METHod:[VALue]?
:<Handle>:SGENeral:PDATa(*):TERMinal(*):OUTPut:OPTic:METHod:[VALue]?
:<Handle>:SGENeral:PPULse(*):TERMinal(*):OUTPut:OPTic:METHod:
[VALue]?
Return Value
Returns the method used to set the optical levels:
•
AVER_ERAT
(Average Power / Extinction Ratio)
•
AVER_MAMP
(Average Power / Modulation Amplitude)
•
HIGH_LOW
(High Power / Low Power)
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com