:SGENeral:GLOBal:ARM Subsystem
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
179
:SGENeral:GLOBal:ARM
Subsystem
Syntax
:<Handle>:SGENeral:GLOBal:ARM[:SEQuence][:LAYer]
This command subsystem is used to control the system by an external
signal.
The EXTernal :INPut connector at the front panel of the clock module
is used to start the system by a so-called GATE or by a TRIGGER
signal.
Following modes are available:
...:SOUR:IMM
A pulse stream (sequence) is generated immediately
as soon as the system is started by the start command.
...:SOUR:GATEd
A pulse stream (sequence) is generated as soon as
the level at the external input connector exceeds the input threshold.
A sequence can be generated by a positive or a by a negative input
signal depending on the parameter set.
...:SOUR:STARtsignal
A pulse stream (sequence) is generated as
soon as a start signal exceeds the input threshold. A sequence can be
started by a positive or a negative input signal depending on the
parameter set.
Pulse stream
Input signal
Pulse stream
Input signal
Pulse stream
Input signal
Pulse stream
Input signal