Output Parameter Commands
Commands in DSR Application Subsystems
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
277
:OUTPut:CAConfiguration[:MODE]
Syntax
:<Handle>:SGENeral:
PDATa(*):OUTPut:
CAConfiguration[:MODE] <OFF | D2
| D4 | A2>
:<Handle>:SGENeral:
PDATa(*):TERMinal(*):OUTPut:
CAConfiguration[:
MODE] <OFF | D2 | D4 | A2>
:<Handle>:SGENeral:
PPULse(*):OUTPut:
CAConfiguration[:MODE]
<OFF | D2 | D4 | A2>
:<Handle>:SGENeral:
PPULse(*):TERMinal(*):OUTPut:
CAConfiguration[:
MODE] <OFF | D2 | D4 | A2>
:<Handle>[:CGRoup(*)]:MODule(*):
CONNector(*):OUTPut:
CAConfiguration[:
MODE] <OFF | D2 | D4 | A2>
Parameters
<OFF | D2 | D4 | A2>
Selects the channel-add mode of a
port/terminal/ connector. The following modes are available:
• OFF (no channel addition). This is the reset value.
• D2 is a digital add of two channels (XOR). It can be used on the
second and fourth frontend slot of an E4832A module if output
frontends are plugged into this slot and into the one immediately
above.
• D4 is a digital add of four channels (XOR). It can be used on the
fourth frontend slot of an E4832A module if the whole module is
equipped with output frontends.
• A2 is an analog add available only for E4838A frontends. It can be
used on the second and fourth frontend slots of an E4832A module
if an E4838A frontend is fitted there.
If A2 Mode is selected, the parameters specified by the following
commands are additionally available:
–
“:VOLTage[:LEVel][:IMMediate]:CAConfiguration:LOW” on
page 229
–
“:PULSe:TRANsition:CAConfiguration[:LEADing]” on page 197
–
“:PULSe:TRANsition:CAConfiguration:TRAiling” on page 199
Example
:_test:SGEN:PDAT1:OUTP:CAC D2
:_test:SGEN:PDAT1:TERM1:OUTP:CAC:MODE D4
:_test:SGEN:PPUL1:OUTP:CAC D2
:_test:SGEN:PPUL1:TERM1:OUTP:IMP:CAC:MODE D4
:_test:cgr1:mod2:conn4:outp:cac A2