Commands in DSR Application Subsystems
Clock Mode Commands
316
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Clock Mode Commands
N O T E
The clock mode command is only valid for the E4862B data generator
frontends.
The clock mode command is used to specify whether pulse (normal) or
clock (high performance) mode is used. In clock mode, a precision
clock signal is generated,with 50 % duty cycle and very low jitter (given
by the clock module — the E4808A clock module should be used).
:SIGNal:MODE
This command sets the mode of the clock signal of E4862B data
generator frontends. The performance mode can be either
CLOCk
(high-
performance) or
PULSe
(normal).
N O T E
Executing this command while the system is running causes the
system to be restarted.
Syntax
:<HANDLE>
[:CGRoup*]:MODule*:CONNector*:DIGital
:STIMulus:SIGNal:
MODE PULSe | CLOCk
:<HANDLE>:SGENeral:
PDATa*:DIGital:STIMulus
:SIGNal:MODE
PULSe | CLOCk
:<HANDLE>:SGENeral:
PDATa*:TERMinal*:DIGital
:STIMulus:SIGNal:MODE
PULSe | CLOCk
:<HANDLE>:SGENeral:
PPULse*:DIGital
:STIMulus:SIGNal:MODE
PULSe | CLOCk
:<HANDLE>:SGENeral:
PPULse*:TERMinal*
:DIGital:STIMulus:SIGNal:MODE
PULSe | CLOCk
Parameters
PULSe
Pulse mode (normal) is used.
CLOCk
Clock mode (high performance) is used.
Example
:_test:SGEN:PDAT1:DIG:SIGN:STIM:MODE CLOC
:_test:SGEN:PDAT1:TERM1:DIG:STIM:SIGN:MODE CLOCk
:_test:SGEN:PPUL1:DIG:SIGN:MODE CLOCk
:_test:SGEN:PPUL1:TERM1:DIG:STIM:SIGN:MODE CLOCk
:_test:CGRoup1:MODule2:CONNector4:DIGital:SIGNal:MODE CLOCk
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com