Commands in DSR Application Subsystems
:SGENeral:GLOBal:SEQuence Subsystem
166
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:<Handle>:SGENeral:GLOBal:SEQuence:
PCONtrol
Syntax
:<Handle>:SGENeral:GLOBal:SEQuence:PCONtrol <NRf>
Parameters
Value in NRf format.
Range/Value Coupling
May influence the data sequence at runtime.
Reset Value
0
Limits
Valid values are 0 and 1.
Description
Sets the value of the program control event. With this event, which is
part of the events defined by the events-expression, the data sequence
can be influenced at runtime. This may be used for “Stop and Go”
sequence model, etc.
Example
SGEN:GLOB:SEQ:PCON 1
:<Handle>:SGENeral:GLOBal:SEQuence:
PCONtrol?
Syntax
:<Handle>:SGENeral:GLOBal:SEQuence:PCONtrol?
Reset Value
0
Description
Returns the state of the program control event line.
Example
SGEN:GLOB:SEQ:PCON?
might return:
1
:<Handle>:SGENeral:GLOBal:SEQuence:
STATe?
Syntax
:<Handle>:SGENeral:GLOBal:SEQuence:STATe?
Description
The sequence indicator returns the number of the currently executed
block. If “0” is returned, the system is stopped or none of the modules
in the system supports the sequence indicator. The sequence indicator
is supported by the modules 3.3G, 7G and 13G.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com