Lightsheet Z.1
Left Tool Area and Hardware Control Tools
Carl Zeiss
02/2013
000000-1790-528
177
It is helpful, especially after an Automatic Detector Alignment has been performed, to use the Profile
View tab in order to evaluate if the image of Cam 1 and Cam 2 overlay. While the grating is continuously
imaged, press the Profile View tab and draw a line or an arrow poly-line on the grating (Fig. 275). Use the
sliders or the input box with arrows of
Cam 1 X
and
Cam 1 Y
(for Channel 1, Cam1) or
Cam 2 X
and
Cam 2 Y
(for Channel 2, Cam 2) to move the lines of the grating to overlay each other.
Leave the Manual Detector Alignment by pressing the
Close
button at the bottom of the window.
Fig. 275
Manual Detector Alignment, Grating, Profile view tab