Maintenance and Cleaning
Lightsheet Z.1
4
000000-1790-528
02/2013
c) To clean and disinfect the sample chamber proceed as follows:
•
Remove the sample chamber and dismantle it into individual parts. Different procedures are required
for cleaning and disinfecting the various parts of the sample chamber.
−
The cover glasses must be properly disposed of in the glass waste.
−
The steel parts of the chamber must be cleaned and disinfected as follows:
Rinse the parts with distilled water.
Lay the parts in an ultrasonic bath containing distilled water to which a few drops of detergent
have been added.
Then rinse the steel parts thoroughly with distilled water until all detergent residue has been
removed.
The parts may be laid in a disinfecting bath of 70
% ethanol (>
1 hour) and then thoroughly
rinsed with distilled water.
The steel parts of the sample chamber may also be autoclaved as necessary.
−
Sealing rings, hose connectors and blank plugs must be cleaned and disinfected as follows:
These parts may not be autoclaved.
First of all lay these parts in distilled water and then in a 70
% ethanol solution for max. one
hour. Rinse them thoroughly with distilled water.
d) To clean and disinfect the temperature sensor proceed as follows:
•
Remove the temperature sensor from the sample chamber and loosen the connections to the main
system module Lightsheet Z.1.
•
Once removed, thoroughly rinse the temperature sensor with distilled water and then suspend it for
one hour in a 70
% ethanol solution. Make sure that the contact plug remains dry. Complete the
cleaning process by rinsing thoroughly with distilled water.