Carl Zeiss
Lightsheet Z.1
4
000000-1790-528
02/2013
•
Use the
Specimen Navigator
to position the capillary from the
Load
position into the
Work
position.
•
Click
Specimen Navigator
.
The
Specimen Navigator
is a graphical tool for accessing the relative positions of the
specimen
embedded in agarose (light-grey cylinder), the
illumination light-sheet
(light-blue), as well as the
X-axis
(red),
Y-axis
(green),
Z-axis
(blue) and the
Rotation
of the sample holder.
Instead of using the
Specimen Navigator,
you
can also move and control the capillary containing
the sample using the
ErgoDrive operating panel
.
Rotary controls:
−
X-axis (red) control
−
Y-axis (green) control (after pressing
Axis
button) Or
Rotary
control (after pressing
Rotation
button)
−
Z-axis (blue) control
Buttons:
The
Rotations
button changes the upper rotary control to move the
rotation
drive.
The
Mode
button changes the
speed
by toggling the drive between coarse and fine.
The
Axis
button changes the upper rotary control to move the
Y drive
.
•
First of all, move the
Y-axis
(green) until the capillary becomes apparent in the image display (from
Load
position to
Work
position).