n
Here you can specify the rated output in 0.0001mV/V. Information to the rated output
can be found in the data sheet of you force transducer.
Byte
Bit 7 ... 0
0...1
Zero balance in 0.0001mV/V
n
Here you can specify the zero balance as 0.0001mV/V value. Information to the zero
balance can be found in the data sheet of you force transducer.
Byte
Bit 7 ... 0
0...1
Gain for user scaling of the output value
n
Here you can specify a factor as 2
-12
value. The factor is multiplied with the output
value.
Byte
Bit 7 ... 0
0...1
User offset for the output value
n
Here you can specify an offset as 2
-12
value. The offset is added to the determined
output value.
Byte
Bit 7 ... 0
0...1
Nominal load of the force transducer
n
Here you can specify the nominal load of the force transducer unit-free. Information to
nominal load can be found in the data sheet of you force transducer.
Byte
Bit 7 ... 0
0...1
Scale factor for the nominal load
n
Here you can specify the scale factor for the nominal load, such as to convert kg to g.
–
Example: Nominal load in kg and scale factor 1000 (03E8h) results display in g.
Byte
Bit 7 ... 0
0...1
Tolerance for
Steady State
n
Here you can specify a tolerance window for the state
Steady State
. This is specified
as a deviation of the scaled nominal load.
–
Example: With a rated load in kg and scaling factor of 1000 (03E8h) you must
specify the value 0005h to set a tolerance window of 5g.
Byte
Bit 7 ... 0
0...1
Time interval for
Steady State
in ms
ZB Zero balance
GN Gain
TA Tara
NL Nominal load
SF Scale factor
SST Steady state toler-
ance
SSW Steady state window
VIPA System SLIO
Analog Input
031-1CA20 - AI 1x16(24)Bit Strain gauge (DMS) > Parameter data
HB300 | SM-AIO | | en | 17-05
184