Installing the Disposables
4-6
20077/021 US
X
TRA
Operator’s Manual
Step 3: Setup of the Vacuum Line
1.
Attach one end of the vacuum line to the X
VAC
system (or another vacuum source) and the other
to the vacuum port (yellow capped) on the lid of the reservoir (see Figure 4-4).
Figure 4-4 Positioning the Blood Collection Reservoir on the Reservoir Holder
If you want to start aspirating from the field:
2.
Make sure the power switch on the back of the X
VAC
is on ON.
Note:
Remove any USB sticks from the rear panel before switching X
TRA
on. The presence of a USB
stick may slow down the booting phase.
3.
Turn on the X
TRA
Note:
When the X
TRA
system is booting up or a fatal error occurred on the X
TRA
system, the X
VAC
vacuum module can be controlled using its built-in control panel (see Figure 3-21). In such cases you
can turn on and control the level of aspiration directly from the X
VAC
control panel (refer to
for detailed information).
Once the machine is ready, the X
VAC
will be controlled only from the X
TRA
screen.
4.
Regulate the vacuum at a level no greater (in absolute value) than 150 mmHg (20 kPa) in
accordance with the AABB guidelines.
5.
Verify that the protective cover on the end of the suction line in the sterile field has been removed
and that the aspiration tip is open to the atmosphere (not blocked).
6.
With the X
VAC
system ON, open the roller clamp on the small bore tubing and allow 200 ml of
anticoagulant solution to be drawn into the Blood Collection Reservoir. This volume should ensure
adequate wetting of the blood contact surfaces. Close the roller clamp if there will be any delay
before processing.
CAUTION
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...