Troubleshooting
X
TRA
Operator’s Manual
20077/021 US
14-5
Memory not functioning.
Part of the data has
been lost.
Warning
Memory might have lost
user protocol names and/or
user protocol parameters.
The machine will continue
working anyway, but it is
recommended to call the
authorized technical
service.
Press X to clear the warning
message and continue
Memory not functioning
Empty clamp test failed
Warning
A hardware anomaly was
detected while testing the
Empty clamp.
Empty clamp test failed
Fill clamp test failed
Warning
A hardware anomaly was
detected while testing the
Fill clamp.
Fill clamp test failed
HCT indicator
calibration failed
Warning
The machine cannot
calibrate the HCT indicator.
• Verify that the line is
properly seated in the
indicator and not dirty on
the outside. Press PLAY to
retry (max. 3 attempts)
• After 3 failures, press
DELAY CALIBRATION to
exit. The system can still
be operated, but the HCT
indicator will not be
available for the current
case.
If the problem persists, call
the authorized technical
service.
HCT indicator calib. failed
Message Text
Warning
or Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management Text
Table 14-1 List of Warnings Which Might Occur During Setup (Continued)
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...