Programmability Option
X
TRA
Operator’s Manual
20077/021 US
9-1
Chapter 9: Programmability Option
Description
The X
TRA
system ships with seven factory-programmed protocols: four ATS protocols (Pstd, Popt, Pfat
and Post-op) and three SEQ protocols (PPP, PRP1, PRP2). The parameters of these factory protocols
have been extensively tested and optimized to produce consistent results. However, by selecting the
Protocol/Mode tab of the Menu Screen (see Figure 9-3), you may create custom protocols which have
different parameters.
It is also possible to change the parameters of factory protocols during a case without creating a
custom protocol. The modified parameters of a factory protocol are stored for only the duration of the
case. Refer to “Modifications Performed in the Protocol/Mode Tab” on page 9-5.
Before creating or editing custom protocols, it is recommended that the operator become familiar with
the factory protocols by reading “Factory Protocols” starting on page 5-22 and “The Preoperative
Sequestration Factory Protocols” starting on page 10-21. Also, to understand the whole blood
separation process (PPP and PRP sequestration protocols), refer to “Overview of Preoperative
Sequestration” starting on page 10-2.
Enabling Programmability Option
In order to create, rename, edit, or delete custom protocols, the “User Protocols” setting must be
enabled from the Protocol Set tab of the Configuration Mode Screen. Refer to “How to Enter the
Configuration Mode” on page 8-2 for instructions on entering the Configuration Mode Screen. For
more on configuring the X
TRA
.
Creating Custom Protocols
Custom protocols are created, modified, and deleted from the Protocol/Mode tab of the Menu Screen
(see Figure 9-3). All instructions in this section assume the Protocol/Mode tab is the currently
displayed screen. Follow these steps to enter the Protocol/Mode tab.
1.
Touch the Menu button from the Setup Screen or the Ready Screen. Protocols cannot be
created, renamed, or deleted during processing.
Note:
Accessing the Protocol tab from the Setup Screen allows the operator to create and modify
protocols for all bowl sizes, while accessing the Protocol/Mode tab from the Ready Screen allows the
operator to create and modify protocols for the current bowl size alone.
2.
From the left side of the Menu Screen, touch the Protocol/Mode tab.
The programmability options of the Protocol/Mode tab are described in “The Protocol/Mode Tab” on
page 9-7.
Entering Values and Text
Before creating or modifying a custom protocol, you should familiarize yourself with the controls for
making modifications.
Numerical values (such as the Fill pump speed) are contained in editable text boxes, and may be
modified in the usual way by touching the text box and using the up and down arrow buttons which
appear next to it (for complete instructions, review “Modification of Parameters” on page 5-5).
CAUTION
The operator is responsible for setting safe parameters for the custom
protocols and for the factory protocols modified during a case.
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...