Troubleshooting
X
TRA
Operator’s Manual
20077/021 US
14-21
Unload limit reached
Warning
You have reached the unload
limit.
1. Remove the pump loop
manually.
2. Switch OFF the machine.
Unload limit reached
Vacuum pump out of
order
Alarm
The vacuum pump is out of
order and has been disabled.
Use an alternative external
vacuum source to continue the
case.
Press V to exit and resume from
the current screen.
WARNING: In POST mode,
proper suction of patient's
drainage blood could be
compromised.
If the problem persists, call the
authorized technical service.
Vacuum pump out of
order
Waste bag full.
Empty or replace it.
Alarm
The volume counter detected
that the waste bag is full.
• If you want to replace it,
press NEW BAG to resume.
WARNING: The waste bag
should be replaced with the
equipment STOPPED and the
bowl completely empty.
• If you want to use the same
bag, empty it to the minimum
level, using the drain at the
bottom of the bag. Once
done, press MINIMUM BAG
LEVEL to resume. This applies
also in case this warning
message appears during
processing.
WARNING: Do not empty
completely the waste bag until
the end of the case. If you
empty the waste bag during the
procedure, leave the minimum
level into the waste bag (see
marking on the waste bag) to
prevent the possibility of
vacuum being generated in the
waste bag during the Empty
cycle. Vacuum in the waste bag
may prevent complete
emptying of the bowl.
Waste bag full
Message Text
Warning or
Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management
Text
Table 14-2 List of Warnings and Alarms Which Might Occur During Processing (Continued)
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...