Overview
2-4
20077/021 US
X
TRA
Operator’s Manual
Reinfusion
Reinfusion of the red blood cells processed by the X
TRA
is under the control and supervision of the
physician in charge and can be accomplished by several different means:
• Processed blood can be held in the primary RBC bag until the bag becomes full. Then the bag can
be disconnected from the autologous transfusion circuit, hung, and reinfused according to normal
blood administration protocol, as described in AABB standards.
• Some of the processed blood can be moved into a transfer pack and handed off to the
anesthesiologist for reinfusion while processing of shed blood continues.
• Blood can be stored for later reinfusion. In this case, follow your institution's blood administration
protocol for appropriate labeling and blood storage requirements.
The Basic X
TRA
The X
TRA
is equipped with an electronic red cell detector which determines when the appropriate
quantity of red cells has accumulated in the centrifuge bowl. This offers three modes for advancing the
blood processing phases:
•
Automatic
: The X
TRA
will fill, wash, and empty the centrifuge bowl continuously until there is no
fluid left to process in the reservoir or until the wash solution is depleted or another waning
condition happens.
•
1 Touch
: The X
TRA
will start automatically when a certain level of fluid is reached in the reservoir,
and will fill, wash, and empty the centrifuge bowl continuously until there is no fluid left to
process in the reservoir or until the wash solution is depleted or another waning/alarm condition
happens.
•
Manual
: The operator has control over each phase transition: from Fill to Wash, and from Wash
to Empty. Options are available to prompt the operator when it is time to proceed to the Wash
phase, and to automatically proceed to the Empty phase.
WARNING
The American Association of Blood Banks provides recommendations
for expiration of salvaged blood (see Warning #11 on page 1-4).
WARNING
Do not reinfuse the patient’s blood from the primary RBC bag when
it is connected to the X
autologous transfusion circuit. Reinfusion
from the primary reinfusion bag connected to the circuit could lead
to air embolism.
WARNING
WARNING
WARNING
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...