Troubleshooting
14-18
20077/021 US
X
TRA
Operator’s Manual
Minimum wash
quality might be not
reached yet
(when processing
with Pfat protocol)
Warning
Based on waste line color
sensor and supernatant
removal algorithm, the wash
quality might not be yet
acceptable.
• Press WASH to clear the
warning message and display
the Wash screen.
• If you want to manually
switch to the Fat Removal
phase, press FAT REM.
• Otherwise, the device will
continue to wash for an
additional volume
corresponding to 50% of the
preset wash volume and
automatically switch to the
Fat Removal phase if
compliant.
Min. quality not reached
PPP bag full. Empty
or replace it.
Alarm
The volume transferred to the
bag exceeds the pre-set safety
level.
Empty or replace the PPP bag,
then press PLAY to resume. The
bag volume counter will be
reset to Zero.
WARNING: Not following these
instructions may expose the
operator to biohazard risk and
compromise the current case.
PPP bag full
Process quickly to
avoid vacuum
deactivation
Warning
The volume of blood inside the
reservoir has reached the
preset safety level.
Process the blood as soon as
possible to avoid vacuum
deactivation.
Process quickly
PRP bag full. Empty
or replace it.
Alarm
The volume transferred to the
bag exceeds the pre-set safety
level.
Empty or replace the PRP bag,
then press PLAY to resume. The
bag volume counter will be
reset to Zero.
WARNING: Not following these
instructions may expose the
operator to biohazard risk and
compromise the current case.
PRP bag full
Message Text
Warning or
Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management
Text
Table 14-2 List of Warnings and Alarms Which Might Occur During Processing (Continued)
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...