Troubleshooting
14-20
20077/021 US
X
TRA
Operator’s Manual
RBCs detected
during purge
Warning
The sensor detected more than
3 ml of RBCs returning into the
bowl.
If you forgot to turn the RBC
bag upside down, press X to
display the Purge screen, then
carefully follow the instructions
displayed on screen.
RBCs during purge
Reservoir empty.
Bowl not filled
Warning
The reservoir might be empty.
The bowl is not properly filled.
• If the reservoir is not empty,
check the air sensor, make
sure the tubing going to the
bowl is seated properly inside
the sensor and the fill line is
not occluded/clamped. Press
FILL to resume.
• If this is the last cycle, press
LAST BOWL (only if enabled)
to automatically complete the
case.
• Press CONCENTRATE to start
the concentration phase.
• Press RETURN (only if
present) to return the blood
into the reservoir.
• If the bowl is sufficiently
packed with RBCs, you may
decide to press WASH to
proceed to the Wash phase or
2XWASH to use twice the
saline solution.
WARNING: Inadequate filling
may lead to lower than
expected hematocrit and
removal of waste components.
Reservoir empty
Reservoir full:
on-board vacuum
pump disabled
Alarm
The vacuum pump has
automatically stopped to
prevent fluid spillage and
serious damage to the pump.
Process the blood immediately.
Once the reservoir level lowers
under the safety level, the
vacuum pump will automatically
reactivate itself.
Res.full:vacuum disabled
Standby too long:
“alarm name”
Warning
The current warning message
has occurred and the centrifuge
has automatically stopped after
5 minutes.
• Press PLAY to resume from
the warning message screen.
• Otherwise, press X to display
the ready screen.
Stand-by too long
Message Text
Warning or
Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management
Text
Table 14-2 List of Warnings and Alarms Which Might Occur During Processing (Continued)
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...