Vacuum Module
X
TRA
Operator’s Manual
20077/021 US
13-3
Preliminary Setup
To use the X
VAC
, attach the vacuum line from the blood collection reservoir vacuum port to the inlet
port on the lid of the vacuum overflow trap. Refer to “Step 3: Setup of the Vacuum Line” on page 4-6.
for information about removing the vacuum overflow trap for
cleaning.
Figure 13-2 Vacuum Line Connecting Reservoir to Inlet Port
Performing a Blood Collection Procedure With the X
VAC
Module
X
VAC
Used Through the X
TRA
System Touch Screen Interface (Remote
Control)
The X
VAC
vacuum module may be used in combination with the X
TRA
system when it is installed into
the X
TRA
cart and is connected to the X
TRA
via the power and serial communications cables:
1.
Connect the X
VAC
to the X
TRA
power supply (see Figure 13-3).
2.
Connect the X
VAC
RS422 female serial connector to the X
TRA
RS422 female serial connector using
the DB9 male to DB9 male cable (see Figure 13-3).
1
1 Vacuum port (yellow capped)
2 Vacuum inlet port
2
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...