Maintenance
X
TRA
Operator’s Manual
20077/021 US
15-5
The trap should be removed and disassembled for cleaning. Push up on the quick-lock connector to
remove the trap from the vacuum system (see Figure 15-4).
Figure 15-4 Removal of the Overflow Trap from the Vacuum Pump
The following drawing illustrates the disassembled parts of the overflow trap.
Figure 15-5 Composition of the Overflow Trap
After removing the trap from the X
VAC
do the following:
1.
Unscrew the jar from the cap.
2.
Remove the float from the jar; pull the rubber splash deflector from the base of the cap.
3.
Clean all surfaces and air passages thoroughly; replace any worn or damaged parts.
4.
Sparingly lubricate the rubber O-ring on the cap with Dow-Corning 111 silicone grease.
5.
Reassemble the trap and reconnect to the X
VAC
Vacuum System.
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...