
Vacuum Module
13-2
20077/021 US
X
TRA
Operator’s Manual
Operating Modes
The operating mode may only be changed when the vacuum pump is off. There are three ways to
change the vacuum operating mode:
• Changing the current active protocol. This is the usual way of changing operating modes. Protocol
changes requiring changes to the vacuum operating mode while the vacuum pump is ON will
produce a warning and prompt the operator to confirm the stop of the vacuum before proceeding.
• Through the Settings tab of the Menu Screen (accessed from the Setup Screen) before the pump
loop has been loaded.
• An “Inconsistent vacuum mode” warning occurs during setup. The operator will be given the
opportunity to change the vacuum mode.
Intraoperative Mode
In the intraoperative operating mode, the vacuum module may be set to provide negative pressure
between -50 mmHg (-6.6 kPa) and -300 mmHg (-40 kPa), with steps of -10 mmHg (-1.3 kPa).
Postoperative Mode
In the postoperative operating mode, the vacuum module may be set to provide negative pressure
between -100 mmHg (-13.3 kPa) and -10 mmHg (-1.3 kPa), with steps of -10 mmHg (-1.3 kPa).
Setting Wakeup Values for Vacuum Module
The operating mode and pressure set points that are used by default when the X
VAC
system is
powered on are determined by how the module was last used.
The wakeup operating mode and setpoint values are determined as follows:
• If the X
TRA
system has no wakeup protocol set, or if it was powered down in the middle of a case,
then the X
VAC
operating mode and setpoints retain the same values they had when the system
was last shutdown.
• If the X
TRA
system does have a wakeup protocol set and no case was interrupted by the last
shutdown, then the vacuum module’s operating mode is set to the X
TRA
wakeup operating mode
(Intra or Post) and the pressure setpoints are set to the corresponding wakeup setpoints (see
below).
All wakeup protocols that provide the Intra aspiration mode have the same Intra wakeup set point.
Likewise, all wakeup protocols that provide the Post aspiration mode have the same Post setpoint.
Follow these steps to set the wakeup vacuum setpoints for each mode:
1.
Enter the Configuration Mode Screen by following the instructions in “How to Enter the
Configuration Mode” on page 8-2.
2.
From the left side of the Configuration Mode Screen, touch the Wake up tab.
3.
From the Wake up tab of the Configuration Mode Screen, locate the “Vacuum Levels” section and
the text boxes labeled “Intra” and “Post-op” at the bottom of the screen.
Modify the value of these boxes to set the wakeup value for the associated mode.
4.
Touch the Close button to return to the Menu Screen. Touch the Close button again to
close the Menu Screen.
For more on configuring the X
TRA
system refer to
.
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...