
Processing
5-18
20077/021 US
X
TRA
Operator’s Manual
Figure 5-23 Reservoir Empty Warning
The operator can proceed in different ways, depending on the following conditions:
Note:
After three minutes, the pump and centrifuge will stop and the Reservoir Empty alarm
with all its buttons will remain displayed.
• If the reservoir is empty and more blood is expected in the reservoir:
a. Wait until the volume of blood collected in the reservoir is sufficient to complete the bowl
filling.
b.
Touch the Fill button to continue with the Fill phase.
• If blood is still present in the reservoir:
a. Check that the connections are properly sealed and eliminate possible occlusions along the fill
line.
Note:
To check the current positioning of the tubing going to the bowl, you may need to press
the Stop button then open the centrifuge and clamp lid. In this case, pay close attention to the
correct disposable installation when closing the clamps and centrifuge lid again.
b. Check that the tube going to the bowl is fully down in the air detector seat.
c.
Touch the Fill button to continue.
• If the reservoir is empty and no more blood is expected in the reservoir:
a. If the RBC bag contains red blood cells:
•
Initiate the Last Bowl function if available (refer to “Last Bowl” on page 7-4).
Note:
The “Last Bowl” function is not available when operating in Manual mode.
•
Otherwise, touch the “Conc” button to complete the bowl filling by pumping the
contents of the RBC bag into the bowl according to the Concentrate function (refer to
“Adding Fluid to a Partially Filled Bowl: The Concentrate Cycle” on page 6-3).
b. If the RBC bag is empty:
•
Either touch the “Return” button to return the contents of the bowl to the reservoir
and then end the case,
•
Or continue processing the partially filled bowl by touching the Wash phase
button (or the “Double Wash” button to use double the wash volume) to advance to the
Wash phase. This will result in lower than usual hematocrit and wash quality of the
collected blood.
Note:
The “Double Wash” button is not available when operating in Manual mode.
Should the message persist, please contact LivaNova Technical Services.
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...