Troubleshooting
14-14
20077/021 US
X
TRA
Operator’s Manual
Centrifuge lid lock
blocked in closed
position
Warning
The centrifuge lid lock is still
closed. Do not try to force
opening.
• Press X to exit and display the
ready screen.
• Otherwise, switching the
machine off and on again
might solve the problem.
If the problem persists, call the
authorized technical service.
Centr. lid lock blocked
Centrifuge lid open
Warning
The centrifuge lid might be
open.
Close the centrifuge lid and
press PLAY to resume.
If the problem persists, contact
the authorized technical
service.
Centrifuge lid open
Check Centrifuge lid
lock
Warning
The centrifuge lid lock is
malfunctioning.
Press X to restore Lock position,
then close the lid.
If the problem persists, call the
authorized technical service.
Check centr. lid lock
Close PPP & Waste.
Open PRP. Then,
press PLAY
Instructional
Message
Follow the on-screen
instructions. Refer to
10: Preoperative Sequestration
.
Close PPP line, open
PRP line. Then, press
PLAY
Instructional
Message
Follow the on-screen
instructions. Refer to
10: Preoperative Sequestration
.
Close PRP & Waste.
Open PPP. Then,
press PLAY
Instructional
Message
Follow the on-screen
instructions. Refer to
10: Preoperative Sequestration
.
Close PRP line, open
PPP line. Then, press
PLAY
Instructional
Message
Follow the on-screen
instructions. Refer to
10: Preoperative Sequestration
.
Do you confirm
empty?
Alarm
This operation may send air
into the RBC bag.
• Press EMPTY to start the
Empty phase.
• Press X to exit and display the
ready screen.
WARNING: Repeating the bowl
emptying several times may
lead to elimination of all
residual air and possible
collapsing of tubings and waste
bag, and inflation of the RBC
bag.
Do you confirm empty?
Message Text
Warning or
Alarm
Buttons Available in the
Button Bar During
Warning/Alarm
Operator Response
Data Management
Text
Table 14-2 List of Warnings and Alarms Which Might Occur During Processing (Continued)
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...