Automated Functions
7-2
20077/021 US
X
TRA
Operator’s Manual
This automation is available only for intraoperative and postoperative protocols (Pstd, Popt, Pfat,
Post-op, and derivatives) in the Automatic and 1 Touch operating modes. It is neither available if the
reservoir type is set to “none” in the Settings tab of the Menu Screen, nor when processing with the
Manual operating mode.
The Autostart function can be activated and deactivated from Protocol/Mode tab of the Menu Screen
by following these steps:
1.
Touch the Menu button from the Ready Screen or any Processing Screen to enter the Menu
Screen.
2.
From the left side of the Menu Screen, touch the Protocol/Mode tab.
3.
From the Protocol/Mode tab, ensure that Automatic or 1 Touch mode is selected from the
mode drop-down menu. Then, touch the Autostart button to toggle the activation of the Autostart
automation. When active, the button appears inset and green.
4.
Optionally, using the text box immediately below the Autostart button, set the Autostart target
volume (ml). This is the volume of fluid which the reservoir must contain before the Autostart
function initiates the cycle.
5.
Touch the Save Modifications button to make all changes in the Protocol/Mode tab take
effect. If the screen is exited by touching the Close button, or is interrupted by an alarm, then the
modifications are not saved.
Refer also to “Set the Active Operating Mode” on page 5-35.
When the Autostart function is active, the Autostart symbol (
) appears in the status area of the
Ready Screen and every processing screen.
The Autostart target volume may be set from the Protocol/Mode tab (as described in the steps above)
or from the Reservoir displet in the Ready Screen and any processing screen. The target volume may
be set from 200 ml - 3,500 ml, in 100 ml steps.
Each bowl size may have a unique Autostart target wakeup volume. Refer to “Setting the Wakeup
Configuration” on page 8-7.
When the system registers the correct conditions for starting the procedure (the reservoir having been
detected to contain at least the Autostart target volume for at least 3 seconds), it displays the
Autostart Activation Screen (see Figure 7-1).
Figure 7-1 Autostart Activation Screen
Содержание 75220
Страница 1: ...Operator s Manual SOFTWARE VERSION 2 00 20077 021 US FOR US MARKET ONLY...
Страница 70: ...System Description 3 32 20077 021 US XTRA Operator s Manual...
Страница 148: ...Special Cycles 6 10 20077 021 US XTRA Operator s Manual...
Страница 156: ...Automated Functions 7 8 20077 021 US XTRA Operator s Manual...
Страница 178: ...Programmability Option 9 10 20077 021 US XTRA Operator s Manual...
Страница 206: ...Preoperative Sequestration PPP and PRP 10 28 20077 021 US XTRA Operator s Manual...
Страница 232: ...Vacuum Module 13 10 20077 021 US XTRA Operator s Manual...
Страница 260: ...Troubleshooting 14 28 20077 021 US XTRA Operator s Manual...
Страница 270: ...Maintenance 15 10 20077 021 US XTRA Operator s Manual...
Страница 282: ...A 6 20077 021 US XTRA Operator s Manual...
Страница 284: ...B 2 20077 021 US XTRA Operator s Manual...
Страница 286: ...C 2 20077 021 US XTRA Operator s Manual...
Страница 298: ...D 2 20077 021 US XTRA Operator s Manual...
Страница 302: ...E 4 20077 021 US XTRA Operator s Manual...
Страница 304: ...LivaNova Deutschland GmbH Lindberghstrasse 25 D 80939 M nchen Germany Tel 49 0 89 32301 0 Fax 49 0 89 32301 555...